NellcorPuritanBennettE 5.4.10

Compatible Applications

Capsule™ Vitals Stream

Device Manufacturer

Covidien

Supported Devices

Ventilator

  • PB 840

Description of the Release

  • Upgraded to Enhanced Data Model (EDM) to support advanced data integration.
  • Fixed Generic Ventilation Mode and Ventilation Mode Primary variables to be output correctly when the “VC” ventilation mode is active without ventilation options.
  • Fixed Percent of Spontaneous Inspiratory Time variable to output a percentage instead of a unitless ratio with a range of 0 – 1.
  • Removed Expired Mandatory Tidal Volume variable. LowerAlarmLimit and UpperAlarmLimit attributes associated with this variable are received on Expired Mandatory Tidal Volume Upper Alarm Limit and Expired Mandatory Tidal Volume Lower Alarm Limit variables.
  • Corrected the documented Alarm Repetition Interval in Helpfile from 1s to 3s.

Cases addressed by this Release

  • C120808 – PB840 Software Version Compatibility