Applications compatibles
Capsule™ Vitals Stream
Fabricant de l'appareil
B. Braun Melsungen AG
Appareils pris en charge
Pompe à perfusion
- Infusomat Compact Plus
- Infusomat fmP (IFPC, IFPE, IFPe)
- Infusomat fmS (IFMC, IFME, IFMe)
- Infusomat Space (686)
- Infusomat Espace P (687)
- Perfuseur compact (PLAA)
- Perfusor Compact Plus
- Perfuseur compact S (PLBE)
- Perfuseur fm (PFAD, PFAE, PFAe)
- Espace Perfuseur (688)
Support/station d'accueil pour pompe à perfusion
- Compact Plus Station
- Gestionnaire de fluides (ordinateur fm)
- Station spatiale (SpaceCom)
Description de la version
- Added support for SpaceCom device firmware version L92.
- Added a known issue for SpaceCom device: “Values exported by a pump can be temporarily and erroneously multiplied by 100 after 30 or more hours of continuous operation with SpaceCom and Space WiFi Battery type 1. This issue has been fixed in firmware version L92.”
Cas traités dans ce communiqué
- C109321 – New firmware L92 for SpaceCom SpaceStation (Philips MRC/BBraun GmbH)
- C114016 – VLKH – BBraunBCC DDI support of firmware L92