CardinalA 5.3.1

Applications compatibles

Capsule™ Vitals Stream

Fabricant de l'appareil

CareFusion

Appareils pris en charge

Ventilateur

  • LTV Series 1000
  • LTV Series 1150
  • LTV Series 1200
  • LTV Series 1200 MRI
  • LTV Series 800
  • LTV Series 900
  • LTV Series 950

Description de la version

  • Added variable 584 (Ventilation Mode) to properly export the active ventilation mode.
  • Added support for protocol revision J and firmware version 5.10.
  • Added I:E Ratio variables.
  • Validated new isolated cable.
  • Added Device entry for LTV Series 1200 MRI.
  • Replaced three legacy waveform variables by CDS variables and added Waveform Output Section in the Help File.
  • Corrected supported firmware versions into Help File.
  • Removed PEEP variables from the LTV Series 800, LTV Series 900, LTV Series 950, and LTV Series 1000 devices because they do not support these variables.
  • Corrected the unit code of some variables.
  • Added Nonsignificant flag for four variables. This flag is associated with some specific values (see help file for details).
  • Added DeviceTime attribute to variables 2255 (Device Date/Time), and to waveforms variables.
  • Removed variables 5926 (Visual Status Flags 1) and 5927 (Visual Status Flags 2)
  • Added MeasurementTime attribute to waveforms variables.
  • Added Alarm repetition section in the Variable Detailed Description to facilitate alarm integration, added additional info for generic Alarm variables.

Cas traités dans ce communiqué

  • C72797 – Need help interpreting Cardinal Health LTV 1200 variable 4129
  • C76372 – Carefusion LTV MRI vent feasible?
  • C79648 – Axon Connectivity with LTV1200
  • C83927 – LTV Series 1200 Waveform DDI Output
  • C98384 – LTV 1000 Firmware Question
  • C100172 – LTV 1200
  • C105102 – Need the CardinalA 5.3.1 for the LTV 1200
  • C15506 – Inova – LTV1200 – Not Sending Mode, Sends All Parameters Regardless of Mode
  • C52655 – CardinalA DDI needs Vent mode Variable added
  • C57497 – LTV-1200 Variable 9113
  • C57498 – LTV1200 needs Vent modes Broken out CardinalA DDI
  • C32009 – CardinalA Driver documentation