Applications compatibles
Capsule™ Vitals Stream
Fabricant de l'appareil
Covidien
Appareils pris en charge
Ventilateur
- PB 840
Description de la version
- Mise à niveau vers le modèle de données amélioré (EDM) pour prendre en charge l'intégration avancée des données.
- Fixed Generic Ventilation Mode and Ventilation Mode Primary variables to be output correctly when the “VC” ventilation mode is active without ventilation options.
- Fixed Percent of Spontaneous Inspiratory Time variable to output a percentage instead of a unitless ratio with a range of 0 – 1.
- Removed Expired Mandatory Tidal Volume variable. LowerAlarmLimit and UpperAlarmLimit attributes associated with this variable are received on Expired Mandatory Tidal Volume Upper Alarm Limit and Expired Mandatory Tidal Volume Lower Alarm Limit variables.
- Corrected the documented Alarm Repetition Interval in Helpfile from 1s to 3s.
Cas traités dans ce communiqué
- C120808 – PB840 Software Version Compatibility