S’identifier Contacter
Demander une démonstration
juin 17, 2020
Capsule DDI - Device Driver Interface - MaquetB 5.2.4

MaquetB 5.2.4

Applications compatibles

Flux de données vitales Capsule™

Fabricant de l'appareil

Getinge

Périphériques compatibles

Intra-Aortic Balloon Pump

  • Cardiosave

Description de la version

  • Added waveform support for Cardiosave
  • Added support of protocol revision C.06.
  • Added MeasurementTime attribute to variables 168 and 4485.
  • Corrected datatype of variable 2255 (Device date/Time) from Double to String in the help file.
  • Marked the variable 5903 (IABP Frequency Setting) as NonSignificant when its value is “INVALID”.

Cas traités par ce communiqué

  • C72025 – Clarify the support of CardioSave in Cardiosave Rescue / Cardiosave Hybrid
  • C61788 – wave forms drivers request cardiosave

Retour